beautypg.com

LumaSense Technologies ISR 320 User Manual

Page 42

background image

ISR 320 Manual

Troubleshooting 42

Symptom

Probable Cause

Comments

Incorrect slope adjustments.

See sections 4.2, 5

The measurement is influenced by reflections
of hot machine parts

Use mechanical
accessories to avoid the

influence of the
interfering radiation.

If target emissivity is less than 1, reflections
from nearby hot objects will influence

readings.

Block reflection paths to
see if readings drop to

expected temperature.

Incorrect spectral response for application.
ISR 320 may be “seeing” through target.

See Technicial
Specifications
Consult LumaSense for
analysis.

Electrical noise in lead wires caused by intense

magnetic fields and/or improper selection of
interconnecting cable and/or cable run in
same conduit as AC power lines.

See section 3.1

Lens or window clouding up.

Increase or decrease air

purge rate and/or
investigate other cause.

Target leaving field of view or ISR 320 mount
slipping.

Faulty ISR 320.

Contact LumaSense. See
section 1.5

Improper grounding of cable shield and/or

ISR 320 housing (or cooling jacket).

ISR 320 mounting not secure.

Noisy readings (fast
fluctuations).

Target moves more than 95% out of field of
view. Conveyor type applications will also
cause fluctuations as targets move by the field
of view.

Application may require
Maximum to be stored.

Apply t

clear

settings. See

section 4.5.

Target may be changing temperature quickly.
(Thermocouples do not respond as fast as the

ISR 320).

Speed of response may

be slowed down Apply t

90

settings. See section 4.4

Noisy readings (fast
fluctuations).

Target may be reflecting nearby objects that
have fluctuating temperatures.

Loose connection in wiring.

Electrical noise lead wires caused by intense
magnetic fields, improper selection of inter-
connecting cable.

Noisy readings (fast
fluctuations).

Cable run into in same conduit as AC power
lines.