beautypg.com

9 troubleshooting – LumaSense Technologies ISR 320 User Manual

Page 41

background image

ISR 320 Manual

Troubleshooting 41

9 Troubleshooting

Symptom

Probable Cause

Comments

No analog output even if the

display shows a temperature
above lower range limit

Reversed leads to ISR 320.

The ISR 320 is diode

protected if power leads
are reversed.
See section 3.1

Reversed leads to other instruments in the

current loop.

Open circuit in lines connecting all instruments
in current loop.

Total burden higher than 500 Ohms

Insufficient supply voltage (See wiring
diagram).

Temperature readings are too

low

Target size too small.
Target is not at correct focus distance.

See section 3.3 and 5.10
With small target/close
focus models, aiming may
require X-Y positioning

table for fine adjustment
during aiming.

Incorrect alignment of the pyrometer to the
object.

Realign pyrometer in
1-channel mode (mono)
to achieve the maximum

temperature signal.

Dirty window or lens.

See sections 4.7 and 6.1

Target temperature too low.

Target emissivity much lower than expected.

Even 2-Color instruments
depend on actual target
emissivity to supply

sufficient energy to the
sensor.
See sections 4.1, 4.2, 4.3,
4.7, 6.1

Incorrect slope adjustments.

See sections 4.2, 5

Faulty ISR 320

Contact LumaSense. See

section 1.5

Temperature readings are too
high

Targets hotter than expected. (This often
happens with first time infrared Pyrometer
users).

Verify target temperature
with contact type
method.
(Thermocouples may not
accurately determine
surface temperature due
to “heat sink” effect).