LumaSense Technologies ISR 320 User Manual
Page 36
ISR 320 Manual
Data format UPP
36
Description
Command Parameters
Measuring value
(one-channel and
ratio temperature)
AAek
Answer: SSSSSQQQQQ
2x5 decimal digits (in °C or °F, last digit is 1/10 °C or °F),
SSSSS = one-channel temperature
QQQQQ=ratio temperature
Emissivity for one-
channel temperature
AAemXXXX
AAem
XXXX = 0100 to 1000
=0.100 to 1.000
Answer: DDDD 4 decimial digits 0100 to 1000
Transmittance of
window
AAetXXXX
AAet
XXXX = 0100 to 1000
=0.100 to 1.000
Answer: DDDD 4 decimial digits 0100 to 1000
K= 1 / 2
Emissivity ratio
AAevXXXX
AAev
XXXX=0800 to1250
1 / 2=0.800 to 1.250
Answer: DDDD
4 decimal digits 0800 to 1250
Response Time t
90
AAezX
X=0
to
7 0=min. 1=0.01s 2=0.05s 3=0.25s
4=1.00s 5=3.00s 6=10s
7=30s
Temp Display °C or °F AAfhX
Output: X=0 Display
in
°C
X=1
Display in °F
Device address
AAgaXX
XX=(00 to 97)
00 to 97=regular device addresses
98 = global address without response
99=global address with response
Internal temperature
(read)
AAgt
AAtm
Answer: DDD 3 decimal digits (000 to 098 °C or
032 to 210 °F)
gt=current temp.
tm=maximum temp. (memory)
Operation
mode
AAkaX
X = 0 metall mode
X = 1 mono mode (1 channel)
X = 2 ratio mode (2 channel)
LED
AAlaX
AAla
X=0 off X=1 on
Answer: 1 digit: “0” or “1”
LED aiming light at
Power-on
AAlpX
X = 0 aiming light off at Power-on
X = 1 aiming light on at Power-on
Software simulation
of external clearance
AAlx
Clears maximum storage
Clear Peak Memory
t
clear
AAlzX
X=0 to 8 0= OFF 1=0.01s
2=0.05s
3=0.25s 4=1.0s 5=5.0s
6=25.0s 7=EXTERN 8=AUTO
Basic
range
(read)
AAmb
Answer: XXXXYYYY
2x4 hex-digit for lower and upper range limit (°C or °F)
Sub
range
(read)
AAme
Answer: XXXXYYYY
2x4 hex-digit for lower and upper range limit (°C or °F)
Sub range (set)
1. AAm1XXXXYYYY XXXXYYYY=2x4 hex-digit for lower and upper sub range
limit (°C or °F)
2. AAm2
AAm2 confirms the change (auto reset)