Alarms, Dsplay, Startng/stoppng computaton – Yokogawa µR10000 User Manual
Page 44

IM 04P01B01-01E
1-28
For the procedure to set the functions, see section 1.10, “Function Setup Guide.”
• The Number of Channels Beng Trend Recorded and the Fastest
Recordng Interval
The fastest recording interval varies depending on the number of channels being
trend recorded. Only display or printout is possible on measurement channels or
computation channels that are not trend recorded.
Number of Channels Beng Trend Recorded
Fastest Recordng Interval
6 channels or less
10 s
7 to 12 channels
15 s
13 to 18 channels
20 s
Enabling/Disabling trend recording for each channel: Section 9.9
Prntout
Various printouts can be executed in the same fashion as measurement channels.
The computed values of computation channels can be printed at constant intervals by
using the periodic printout timer or the two TLOG timers. The TLOG timers can be used
only on the computation channels. For the printout contents when printing using the
TLOG timer, see appendix 1.
Setting the printout using the TLOG timer: Sections 9.6 and 9.12
Alarms
Up to 4 alarms can be assigned to each computation channel.
• The alarm types are high limit alarm (H), low limit alarm (L), delay high limit alarm (T),
and delay low limit alarm (t).
• Hysteresis can be specified to the values for activating and releasing the alarm. The
hysteresis applies only to high limit alarm (H) and low limit alarm (L). The hysteresis
width can be set in the range of 0.0% (Off) to 1.0% of the recording span in 0.1
steps. The setting applies to all high limit alarms and low limit alarms. By default, the
hysteresis width is set to 0% (Off).
Setting the alarm delay function: Sections 7.15 and 9.11
Setting the alarm hysteresis: Section 7.1
Dsplay
Computation channels can be displayed on the VFD as with the measurement channels.
Startng/Stoppng Computaton
Computation can be started or stopped using keys.
1.6 Computaton Functon (/M1 Opton)