App index lnear scalng – Yokogawa µR10000 User Manual
Page 103

5-3
IM 04P01B01-01E
Frequently
Used
Setup
Operat
ons
(Sett
ng
Mode)
5.1 Settng the Input Range
1
2
3
4
5
6
7
8
9
10
11
12
App
Index
Lnear Scalng
Select Scale.
Set the channel range.
Select Volt, TC, RTD, or DI.
Select the range type.
Set the left span value.
Set the right span value.
Set the decimal position
and left scale value.
Set the right scale value.
The new setting takes effect.
First channel Last channel
ESC/?
Set=Range
CH=01-01
Mode=Scale
Type=Volt
Range=2V
Span_L= -2.000
Span_R= 2.000
Scale_L= 0.00
Scale_R= 200.00
01-01 Channel
Setting complete
Descrpton
• Scale Left, Scale Rght, and Decmal Poston
Selectable range (mantissa): –20000 to 30000
Example: The value in the range of –100.00 to 350.00 cannot be specified. The
mantissa of Scale_R is 35000, which exceeds the upper limit of 30000. Set
in the range of –100.0 to 350.0.
Decimal position: XXXXX, XXXX.X, XXX.XX, XX.XXX, X.XXXX
The decimal position is set using Scale_L. It cannot be set using Scale_R.
Example: Change 0.00 to 0.0.
Press the
key to move the cursor to the last digit of 0.00.
Space
Press the key to change the value to 0.0 .
Note
The displayable/printable range of scale values is –19999 to 30000 excluding the decimal.