App index – Yokogawa µR10000 User Manual
Page 259

App-7
IM 04P01B01-01E
Append
x
Appendx 2 Perodc Prntout Interval
1
2
3
4
5
6
7
8
9
10
11
12
App
Index
When the Interval Is Set to Manual
If you select Manual, you can set the interval.
• Chart Speeds That Allows Perodc Prntout
Pen Model
Dot Model
Chart Speed
Prntout Avalablty Chart Speed
Prntout Avalablty
Up to 9 mm/h
No
Up to 9 mm/h
No
10 to 1500 mm/h
Yes
10 to 100 mm/h
Yes
1600 mm/h or higher
No
101 mm/h or higher
No
• Interval Settngs
Select from 10, 12, 15, 20, 30 minutes, 1, 2, 3, 4, 6, 8, 12, and 24 hours.
• Prntout Items
Item
Number of Lnes
Intal Settng
Notes (Ref. Secton)
Date/Time
2
-
Always printed.
Measured value and
alarm status
1 line/channel
Printed
Print or not print selectable for
each channel (section 6.6).
Scale
2
Printed
Print or not print selectable
(section 7.7).
Pen color
(Pen model only)
1
Printed
Print or not print selectable
(section 7.7).
Chart speed
1
-
Always printed.
• Relatonshp between the Prntout Items and Interval
The next periodic printout does not start until all items of the previous periodic printout
are printed.
If all items cannot be printed in the specified interval, the printout interval increases by
an integer multiple of the specified interval.
Periodic printout
Periodic printout
Actual
interval
Specified interval
Example: Dot model, chart speed: 20 mm/h, interval: 30 minutes, all other settings at
default.
• This setting requires 11 lines to be printed which is equivalent to
approximately 28 mm* of chart paper.
• The length of chart paper that is fed within the interval is 10 mm.
* The length of chart paper needed in the printout varies depending on the chart
speed and dot printing interval.
Not all items can be printed in the specified interval of 30 minutes.
The actual interval is 1 hour and 30 minutes.
If you want to print at a shorter interval, change the settings as follows:
• Reduce the number of printed items.
• Increase the chart speed within the printable range.
Note
Measurement, display, periodic printout, and trend recording (dot model) are not performed by
setting unused channels to Skip (see section 5.1).