7-5
SM 04L41B01-01E
Replacing Parts
1
2
3
4
5
6
7
5.
Open the bezel upward, then fix in the upright position with the side stay.
Stay
6.
Turn OFF the power switch. Turn ON when reinstalling the inner instrument.
1
0.4
N.
m
Fastening
Torque
3
2
4
Power switch
7.
Loosen the screw. (Screw tightening torque: 0.4 to 0.5 N·m).
1
0.4
N.
m
Fastening
Torque
Screw
USB
CF CARD
3
4
5
8.
Pull out the handles in the order handle 2, handle 3, then handle 4. The handles
pull out approximately 8 mm.
Handle 2
Handle 3
Handle 4
USB
CF CARD
2
3
4
5
1
0.4
N.
m
Fastening
Torque
9.
Pull on handle 2 while pushing the lock release lever to the left, then pull out the
inner instrument.
Lock release lever
USB
CF CARD
1
0.4
N.
m
Fastening
Torque
3
4
5
7.3 Pulling Out the Inner Instrument (DX1000N)