beautypg.com

Yokogawa Removable Chassis DX1000N User Manual

Page 21

background image

3-7

SM 04L41B01-01E

Testing

1

2

3

4

5

6

7

Reference Value (Integral Time: 20 ms (50 Hz)/16.7 ms (60 Hz))

Range

Input Value

Range Allowed

by Test

Accuracy Specification Note

20 mV

+20.000 mV

±18 digits

(0.05% of rdg + 12 digits)

0.000 mV

±10 digits

–20.000 mV

±18 digits

60 mV

+60.00 mV

±5 digits

(0.05% of rdg + 3 digits)

0.00 mV

±2 digits

–60.00 mV

±5 digits

200 mV

+200.00 mV

±10 digits

(0.05% of rdg + 3 digits)

0.00 mV

±2 digits

–200.00 mV

±10 digits

2 V

+2.0000 V

±18 digits

(0.05% of rdg + 12 digits)

0.0000 V

±10 digits

–2.0000 V

±18 digits

6 V

+6.000 V

±5 digits

(0.05% of rdg + 3 digits)

0.000 V

±2 digits

–6.000 V

±5 digits

20 V

+20.000 V

±10 digits

(0.05% of rdg + 3 digits)

0.000 V

±2 digits

–20.000 V

±10 digits

50 V

+50.00 V

±4 digits

(0.05% of rdg + 3 digits)

0.00 V

±2 digits

–50.00 V

±4 digits

Pt100

–200°C: 18.52 Ω

±0.4°C

(0.15% of rdg + 0.3°C)

0°C: 100.00 Ω

±0.2°C

600°C: 313.71 Ω

±0.9°C

TC-T

0°C: 0.000 mV

±0.4°C

(0.15% of rdg + 0.5°C)

EXT RJC mode

TC-K

0°C: 0.000 mV

±0.5°C

(0.15% of rdg + 0.7°C)

EXT RJC mode

Cu10

(GE)

–200°C: 1.326 Ω

±1.4°C

(0.4% of rdg + 1.0°C)

Performed when

the /N1 option is

specified.

0°C: 9.036 Ω

±0.8°C

300°C: 20.601 Ω

±1.7°C

Pt50

–200°C: 8.57 Ω

±0.8°C

(0.3% of rdg + 0.6°C)

Performed when

the /N3 option is

specified.

0°C: 50.00 Ω

±0.4°C

550°C: 150.40 Ω

±1.8°C

Reference Value (Integral Time: 1.67 ms (600 Hz))

Range

Input Value

Range Allowed

by Test

Accuracy Specification

2 V

+2.0000 V

±57 digits

(0.1% of rdg + 40 digits)

0.0000 V

±32 digits

–2.0000 V

±57 digits

* Not tested for the DX2000 with the /MC1 option.

3.4 Test Procedures

This manual is related to the following products: