3-6

SM 04L41B01-01E

Measurement Accuracy

● Overview

The measurement accuracy is tested using two different* integral times.

* But only one integral time is used when using the DX2000 with the /MC1 option.

The measurement accuracy test is performed in one of the following ways.

• One representative channel for each A/D converter is tested, and the channel-to-

channel error between the representative channel and other channels is tested (see
next page).

* Choose any channel number for each A/D from those listed in the A/D converter

configuration table.

Ex.: With the DX1004, set channels 1 and 3 as representative channels.

• The measuring accuracy of all channels is tested separately.

● A/D Converter Configuration

Models

Measurement

Interval

Number of

Channels

No. of

A/Ds

Channel Number by A/D

DX1002/DX1002N

125 ms

2CH

1

1-2

DX1004/DX1004N

125 ms

4CH

2

1-2, 3-4

DX1006/DX1006N

1 s

6CH

1

1-6

DX1012/DX1012N

1 s

12CH

1

1-12

DX2004

125 ms

4CH

2

1-2, 3-4

DX2008

125 ms

8CH

4

1-2, 3-4, 5-6, 7-8

DX2010

1 s

10CH

1

1-10

DX2020

1 s

20CH

2

1-10, 11-20

DX2030

1 s

30CH

3

1-10, 11-20, 21-30

DX2040

1 s

40CH

4

1-10, 11-20, 21-30, 31-40

DX2048

1 s

48CH

4

1-12, 13-24, 25-36, 37-48

● Test Instruments

Instrument

Specifications

DC voltage generator

Accuracy: ±0.005% of setting + 1 µV

DMM

Accuracy: ±0.005% of rdg + 1 µV

Variable resistors

Accuracy: 0.01% + 2 mΩ or better

● Setup

Set the input range of the channels under test, and the test range.

● Test Procedures

Warm up the instrument for thirty minutes before performing the test.
Enter the input value determined for each range. Check whether the displayed

measured values satisfy the reference values.

DC voltage generator

To the input

terminals

+

Variable resistors

The resistance of three

lead wires must be equal.

3.4 Test Procedures