SM 04L41B01-01E
4.2 Adjusting the A/D Converters
● Overview
The number of included A/D converters differs depending on the instrument model.
For all of the A/D converters, input a predetermined value for each range and take
measurements, then save the calibration values.
● Instruments Used
Instrument
Specifications
DC voltage generator
Accuracy: ±0.005% of setting + 1 µV or better
Variable resistors
Accuracy: 0.01% + 2 mΩ or better
● Calibration
Calibrated Channels
Model
A/D No. 1
A/D No. 2
A/D No. 3
A/D No. 4
Zero
Full
Zero
Full
Zero
Full
Zero
Full
DX1002/DX1002N
CH1
CH2
-
-
-
-
-
-
DX1004/DX1004N
CH1
CH2
CH3
CH4
-
-
-
-
DX1006/DX1006N
CH1
CH2
-
-
-
-
-
-
DX1012/DX1012N
CH1
CH2
-
-
-
-
-
-
DX2004
CH1
CH2
CH3
CH4
-
-
-
-
DX2008
CH1
CH2
CH3
CH4
CH5
CH6
CH7
CH8
DX2010
CH1
CH2
-
-
-
-
-
-
DX2020
CH1
CH2
CH11
CH12
-
-
-
-
DX2030
CH1
CH2
CH11
CH12
CH21
CH22
-
-
DX2040
CH1
CH2
CH11
CH12
CH21
CH22
CH31
CH32
DX2048
CH1
CH2
CH13
CH14
CH25
CH26
CH37
CH38
Input Value
Input Type
Range
Zero
Full
Remarks
DC voltage
20 mV
0 mV
20 mV
60 mV
0 mV
60 mV
200 mV
0 mV
200 mV
1 V
0 mV
1 V
2 V
0 mV
2 V
6 V
0 mV
6 V
20 V
0 mV
20 V
50 V
No adjustment needed
RTD
Pt100
0 Ω
300 Ω
Cu10
0 Ω
200 Ω
With the /N1 or /N3 option.
Connection
UUA
When DC range
DC Voltage
generator
Channel 1 (CH1)
H
L
UUA
When RTD or CU range
–
Channel 1 (CH1)
b
DC Voltage
generator
Variable
resistors
Channel 2 (CH2)
H
L
Channel 2 (CH2)
+
– +
B A
b B A
(Zero value)
(Full value)
(Full value)
UUA: Unit Under Adjustment
Short these
three wires.
(Zero value)