beautypg.com

Yokogawa Removable Chassis DX1000N User Manual

Page 26

background image

4-3

SM 04L41B01-01E

Adjustments

1

2

3

4

5

6

7

● Calibration Procedure

1.

Press the

UP arrow key while turning ON* the power.

The instrument starts up in

Calibration Mode, and the A/D No. are displayed.

* The DX1000N power supply switch is located inside the bezel. Disconnect then

reconnect the power supply, or open the bezel and operate the power supply switch (see
section 7.3).

2.

Select the

A/D No. to calibrate with a soft key, then press the DISP/ENTER key.

DX1000/DX1000N

DX2000

3.

Press the

Cal/Exec #2 soft key.

DX1000/DX1000N

DX2000

Shows the day the A/D most

recently adjusted

Inputting a Reference Signal

4.

Input the zero value of the range to be adjusted to the zero channel of the A/D

No., and the full value to the Full channel. See the calibration channels and input
values on the previous page.

Ex.: To adjust the 20 mV range of A/D No. 1, input 0 mV to CH1, and 20 mV to CH2.

Executing the Adjustment

5.

Select the soft key of the range to adjust, then press the

DISP/ENTER key.

DX1000/DX1000N

DX2000

Shows the A/D No. being

calibrated

Select the Range to Calibrate

A message is displayed, and adjustment executes automatically.

M550

The A/D calibration is being

executed...

6.

Repeat steps

4

and

5

to adjust all ranges.

When finished, press the

ESC key.

4.2 Adjusting the A/D Converters

This manual is related to the following products: