2-2

SM 04L41B01-01E

Is the key lock set?

Have measures against

external noise been taken?

See No. 4 of the checklist.

See No. 2, 3, 7, 9, and 10 of the

checklist in section 2.2.

Release the key lock setting.

Keys do not operate correctly

If there is a problem with a display

screen or other function

NO

YES

YES

NO

• Keep measurement input,

communication, and other

wires and cables away from

noise sources.

• Be sure to ground all

instruments.

• Use shielded cables for

wiring measurement input,

communication, and other

circuits.

• Use isolation transformers

on power supply lines.

• Insert filters on power supply

lines.

• If inductive loads are used

on alarm lines, insert surge

absorbers in the lines.

Note

If menu items required for screen menus or FUNC key menus are not displayed, check menu
customization settings.

Is the memory medium

(CF card) working normally?

Is it a recommended

device?

See No. 8 of the checklist

in section 2.2.

Replace the memory medium

(CF card) with a new one.

Malfunction of external medium

(CF card)

YES

NO

2.1 Failure Analysis Flow Chart