2.2 Troubleshooting Checklist

The table below describes the most common types of failures and their corrective
actions.

No. Phenomenon

Action

Description

Refer to Chapter 6

Check

Adjust

Replace

*1

Part Name

DX1000

DX1000N

DX2000

Page Item Page Item Page Item

1

The instrument

doesn’t start

even though

the power is

ON

Power supply cable connection/wiring -

-

-

-

-

-

-

MAIN POWER BOARD ASSEMBLY

Power PBA

6-4

4

6-13

3

6-23

6

MAIN BOARD ASSEMBLY

Main PBA

6-4

12

6-13

14

6-23

18

SUB BOARD ASSEMBLY

Sub PBA
EXP. BOOT

6-4

13

6-13

15

6-23

19

2 FAIL state

MAIN BOARD ASSEMBLY

Main PBA

6-4

12

6-13

14

6-23

18

SUB BOARD ASSEMBLY

Sub PBA
EXP. BOOT

6-4

13

6-13

15

6-23

19

OPTION TERMINAL ASSEMBLY

Option

Terminal

Assembly

6-6

4

6-16

4

6-26

14, 17

20, 23

3

Abnormal

functioning of

memory
(backup)

Battery connection

Battery

Assembly

6-4

14

6-13

16

6-23

20

Battery voltage

(must be +3.0 V or more)

-

-

-

-

-

-

-

MAIN BOARD ASSEMBLY

Main PBA

6-4

12

6-13

14

6-23

18

SUB BOARD ASSEMBLY

Sub PBA
EXP. BOOT

6-4

13

6-13

15

6-23

19

4

Key operation
abnormality

 Check the FFCs

*2

for key wiring

(whether pulled out or damaged)

-

-

-

-

-

-

-

SW & Sensor BOARD ASSEMBLY

Key Case

Assembly

6-3

14

6-12

20

6-22

4

CONNECTION BOARD ASSEMBLY

CONN. PBA

6-3

6

6-12

7

-

-

MAIN BOARD ASSEMBLY

Main PBA

6-4

12

6-13

14

6-23

18

5 LCD display

not normal

 Check the FFCs

*2

for key wiring

(whether pulled out or damaged)

-

-

-

-

-

-

-

MAIN BOARD ASSEMBLY

Main PBA

6-4

12

6-13

14

6-23

18

SUB BOARD ASSEMBLY

Sub PBA
EXP. BOOT

6-4

13

6-13

15

6-23

19

CONNECTION BOARD ASSEMBLY

CONN. PBA

6-3

6

6-12

7

-

-

Back Light Unit of LCD ASSEMBLY

Back Light

Unit

6-3

3

6-12

4

6-22

3

LCD & INVERTER UNIT

LCD

6-3

3

6-12

4

6-22

1

6

Large

measurement

error
Temperature

measurement

abnormal

Keep input wires away from noise

sources (through distance and

shielding, etc.)

-

-

-

-

-

-

-

Check that the input terminals are not

disconnected from the instrument

-

-

-

-

-

-

-

Check that the input terminal cover is

not loose

-

-

-

-

-

-

-

Check that the RJC (INT/EXT) setting

is correct

-

-

-

-

-

-

-

 

AD SCANNER BOARD ASSEMBLY

AD-STD/

ISO

6-4

22

6-13

26

6-23

40, 41

42, 43

7

Fluctuation

in measured

values

Does the integral time setting of the

A/D converter match the power supply

frequency?

-

-

-

-

-

-

-

Keep input wires away from noise

sources (through distance and

shielding, etc.)

-

-

-

-

-

-

-