Communication commands, Specifying the base position of the bar graph – Yokogawa Removable Chassis DX1000N User Manual
Page 19
19
IM 04L41B01-03E
Communication Commands
Communication commands are described in chapter 3, “Commands” in the DX1000/
DX1000N/DX2000 Communication Interface User’s Manual (IM 04L41B01-17E).
Specifying the Base Position of the Bar Graph
SB Sets the bar graph for each channel
Syntax
SB p1,p2,p3
p1 Measurement/computation/external input channel number
p2 Base position of the bar graph display
NORMAL
Normal (lower limit)
CENTER
Center
LOWER
Lower limit
UPPER
Upper limit
p3 Number of scale divisions (4 to 12)
Section 3.4, “Setting Commands (Setting)” in the communication interface user’s manual
Settings on the DX: Page 6
Storage Method for Constantly Retaining the Most Recent Data Files in the CF
Card (Media FIFO)
WU Sets the environment
Set the media
Syntax
WU p1,p2,p3
p1 Setting type (MEDIA)
p2 Auto save function (OFF or ON)
p3 Media FIFO (OFF or ON)
Example Use media FIFO.
WUMEDIA,ON,ON
Section 3.6, “Basic Setting Commands” in the communication interface user’s manual
Settings on the DX: Page 12
Format of the File List Output Using the ME Command
Because the method of assigning the file name has been changed, the output format has
also changed.
The number of output file name characters is 51.
● Syntax
EACRLF
yy/mo/dd_hh:mm:ss_ssssssss_fff•••_0_xxx•••CRLF
••••••••••••••••••••••••••••••••••••••••••••••
ENCRLF
fff••• name (51 characters including the extension. If it is less than 48, spaces
are entered. If this is a directory, the characters
position displaying the file data size.
Section 4.2, “Output Format of ASCII Data” in the communication interface user’s manual