Yokogawa DO202 2-Wire Dissolved Oxygen Analyzer User Manual
Page 72

IM 12J05C01-01E
Appendix 2-7
3/4
QIS 12J05C01-21E
3.5 Galvanic
Indication
Check
Following Section 3.4, press the [ENT] key until the message display shows “GALVA.” Then
turn SW1 off. Change the current of the DC generator 1 and check the data display. The value
on the data display must be within the range shown in Table 3.
Table 3
Reference (μA)
Data Display (μA)
10 10
±0.06
30 30
±0.1
50 50
±0.2
3.6 Polarograph
Indication
Check
Press the [ENT] key until the message display shows “POLAR.1.” Then turn SW1 on. Change
the current of the DC generator 2 and check the data display. The value on the data display
must be within the range shown in Table 4.
Table 4
Reference (nA)
Data Display (nA)
2 2
±0.02
6 6
±0.04
12 12
±0.07
Press the [ENT] key. The message display will show “POLAR.2.” Change the current of the DC
generator 2 and check the data display. The value on the data display must be within the range
shown in Table 5.
Table 5
Reference (nA)
Data Display (nA)
20 20
±0.1
60 60
±0.3
100 100
±0.5
Press the [ENT] key. The message display will show “POLAR.3.” Change the current of the DC
generator 2 and check the data display. The value on the data display must be within the range
shown in Table 6.
Table 6
Reference (nA)
Data Display (nA)
200 200
±1
600 600
±3
1000
1000
±5
Press the [ENT] key until the message display shows “READY.” Press the [ENT] key again to
finish the tests.