2. commissioning mode, 2-1 output range – Yokogawa DO202 2-Wire Dissolved Oxygen Analyzer User Manual
Page 43

IM 12J05C01-01E
5-3 Parameter Setting
5-2-1 Output range
5-2. Commissioning mode
In order to obtain peak performance from the EXA
transmitter, you must set it up for each custom
application. Refer to APPENDIX A-1-2 for stand-
ard configurations and options.
Output range: mA output is set as default
to 0 - 20ppm.
For enhanced resolution in more
stable measuring processes, it may
be desirable to select for example
5-10 ppm range.
Hold
The EXA DO202 transmitter has the
ability to “HOLD” the output during
maintenance periods. This
parameter should be set up to hold
the last measured value, or a fixed
value to suit the process.
Service
This selection provides access to
the service menu.
What follows are pictorial descriptions of typical
frontplate pushbutton sequences for each param-
eter setting function. By following the simple YES/
NO prompts and arrow keys, users can navigate
through the process of setting range, setpoints,
hold, wash and service functions.
mA
mA
mA
NO
NO
NO
YES
ENT
YES
NO
ENT
ENT
ENT
ENT
ENT
ENT
ENT
ENT
YES
NO
YES
NO
YES
NO
YES
Note:
When ppb or %sat
is enabled in code 56,
the output range is set in
a similar way to ppm.
ENT
FAIL
TEMP.MAN
YES NO
ENT
NO
MODE
YES
MODE
%sat
ppb
CAL
DISPLAY
ppm
HOLD
OUTPUT
SET HOLD
SERVICE