6-2 Calibration Procedure

6-2. Calibration procedure using air calibration

method

6-2-1. Preparation

Move the sensor to a maintenance site and wash

off any dirt on the membrane. Lightly wipe off any

remaining water from the membrane with a soft tis-

sue.

Expose the sensor to the air in an environment

where there is no temperature change and no wind

(a convenient way to accomplish this environment is

to place the sensor in an empty bucket).

Note:

If sensors are used without temperature

compensating element, the temperature

must be adjusted to ambient temperature in

service code 13.

6-2-2. Procedure for air calibration

NO

SC21=0

NO

YES

NO

YES

NO

The instruments waits for

the reading to stabilise.

(The display flashes)

When reading Is stable,

the CAL END message

appears.

NO

YES

YES

12J6B2-22

Press the MODE key. CALIB appears in the display, and the

YES/NO

key flags flash. Press YES.

NO

YES

NO

YES

YES

NO

After briefly displaying WAIT,

the display returns to the

normal readout.

YES

Sensor can be returned

to process water.

Press YES for single point

adjustment.

ABLE

NO

NO

YES

NO

NO

YES

YES

Go to

6-4-2

Go to

6-4-3

Go to

6-3-3

NO

NO

YES

SC21=1

NO

ENT

FAIL

TEMP.MAN

YES NO

ENT

NO

MODE

YES

MODE

%sat

ppb

CAL
DISPLAY

ppm

HOLD

OUTPUT

SET HOLD

SERVICE

YES

NO

YES

YES

YES

YES

YES

YES

to *1

NO

from *1

NO

NO

to *1

Note: To start calibration with a zero oxygen

(N2

gas.)