6. PARAMETER DESCRIPTION
T0610-2.EPS
Item
Name
Display unit
(BRAIN)
Data range
Display unit
/BRAIN
Default value
(*): Indicated item
Units
R/W
Position
of decimal
point
Description
J30 Power Synch
(POWER SYNCH)
W
No
Yes
Yes
Selects whether or not the internal
frequency is to be synchronized
with the power supply frequency.
J40 Memo 1
(MEMO 1)
W
ASCII 16 characters
Memo field
J41 Memo 2
(MEMO 2)
W
ASCII 16 characters
Memo field
J42 Memo 3
(MEMO 3)
W
ASCII 16 characters
Memo field
J50 Software Rev No
(SOFTWARE REV)
R
—
Software revision number
J60 —
(SELF CHECK)
R
Good
Error
See “6.5 Alarm Functions”.
J31 Power Frequency
(POWER FREQ)
R/W
Hz
47.00 to 63.00
2
50.00
Displays the power-supply
frequency (for Power Synch =
“Yes”), or sets the power-supply
frequency (for Power Synch=“No”).
(10) Item K (Menu K): Diagnostic Function Setting items
Menu K contains items that are relevant to the diagnosis of insulation adhesion to the electrode.
T0611.EPS
Item
Name
Display unit
(BRAIN)
Data range
Display unit
/BRAIN
Default value
(*): Indicated item
Units
R/W
Position
of decimal
point
Description
K60 —
(SELF CHECK)
R
Good
Error
See “6.5 Alarm Functions”.
K00 Diagnosis
(DIAGNOSIS)
K10 Adhesion Check
(ADHESION CHK)
W
No
Yes
No
Selects whether or not to perform diagnosis
of adhesion to the electrode.
K11 Adhesion Level1
(ADH LEVEL1)
W
M ohm
0.00 to 100.00
2
0.10
Sets the resistance value for adhesion
Level 1 to the electorode.
K12 Adhesion Level2
(ADH LEVEL2)
W
M ohm
0.00 to 100.00
2
0.50
Sets the resistance value for adhesion
Level 2 to the electorode.
K13 Adhesion Level3
(ADH LEVEL3)
W
M ohm
0.00 to 100.00
2
1.00
Sets the resistance value for adhesion
Level 3 to the electorode.
K14 Adhesion Level4
(ADH LEVEL4)
W
M ohm
0.00 to 100.00
2
3.00
Sets the resistance value for adhesion
Level 4 to the electorode.
K15 Adh Measure Value
(ADH MEAS VAL)
R
M ohm
—
2
Displays the resistance value for
adhesion to the electrode.
(11) Item M (Menu M): Automatic Zero-Adjustment Function Setting items
Menu M contains items that are relevant to automatic zero adjustment.
Item
Name
Display unit
(BRAIN)
Data range
Display unit
/BRAIN
Default value
(*): Indicated item
Units
R/W
Position
of decimal
point
Description
T0612.EPS
M60 —
(SELF CHECK)
R
Good
Error
See “6.5 Alarm Functions”.
M11 Magflow Zero
(MAGFLOW ZERO)
M00 Adjustment
(ADJUSTMENT)
R/W
-99.999 to 99.999
3
0.000
Displays the result of the automatic
zero adjustment, or sets the zero point.
M10 Auto Zero Exe
(AUTOZERO EXE)
W
No Execution
Execution
No Execution
Selects whether or not automatic
zero adjustment is carried out.
Linked with B50.