beautypg.com

2 display modes, Display modes -10 – Yokogawa AXFA14G/C User Manual

Page 35

background image

IM 01E20F02-01E

5-10

5. EXPLANATION OF BASIC ITEMS

Arithmetic Out

The display is given in the same manner as Integrator
Out, decimal point is set by
"AR:OUT_RANGE.Decimal_Point".

AR:OUT_RANGE.Decimal_Point

AR:OUT_RANGE.Decimal_Point

Desimal Point

Location

Example

0

0

12345678

1

1

1234567.8

2

2

123456.78

3

12345.678

4

1234.5678

5

123.45678

6

12.345678

7

3
4
5
6
7

1.2345678

Other

0

12345678

T0513.eps

PD Tag

The third line shows the example of PD Tag display.
Maximum number of figures is 16, on LCD the head
16 charactors out of 32 charactors of management PD
Tag can be displayed.




1 2 3

.

4 5

l / h

6

8 7 6 5 4

M l

9

T A G

F R

I T 1

A X F - P l

1

e

c

a

F0511.eps

Adhesion Check

When Adhesion Check display is selected result is
displayed as below showing the level number.

1 2 3

.

4 5

l / h

6

L e

F R

A D H

:

v e l

2

F0512.eps

Adhesion Check Level Judge

Level4 : Adh Measure Value > Adhesion Level4

1 2 3

.

4 5

l / h

6

L e

F R

A D H

:

v e l

4

F0513.eps

Level3 : Adhesion Level4

Ն Adh Measure Value >

Adhesion Level3

1 2 3

.

4 5

l / h

6

L e

F R

A D H

:

v e l

3

F0514.eps

Level2 : Adhesion Level3

Ն Adh Measure Value >

Adhesion Level2

1 2 3

.

4 5

l / h

6

L e

F R

A D H

:

v e l

2

F0515.eps

Level1 : Adhesion Level2

Ն Adh Measure Value >

Adhesion Level1

1 2 3

.

4 5

l / h

6

L e

F R

A D H

:

v e l

1

F0516.eps

Level0 : Adhesion Level1

Ն Adh Measure Value

1 2 3

.

4 5

l / h

6

L e

F R

A D H

:

v e l

0

F0517.eps

Communication

The third line shows the example of Communication
display.

1 2 3

.

4 5

l / h

6

8 7 6 5 4

M l

9

C O M

F R

I T 2

d a t

i o n

F o u n

d b u s

F i e l

F0518.eps

5.6.2 Display Modes

ADMAG AXF has following display modes.

- Normal Display
- Alarm Display
- Warning Display
- Autozero Display

Display Renewal Time

Display renewal time for each display modes depends
on Display Cycle setting;

2s,1s,400ms,200ms

4s

8s

Normal Display

2s

4s

8s

Alarm Display

4s

4s

8s

Warning Display

4s

4s

8s

Autozero Display

Display Cycle set Time

30s

Display Mode

T0514.eps

Normal Display

In this display mode various flow data from one to
three lines are displayed as described in 5.5.1.

This manual is related to the following products: