beautypg.com

4 dimensions, Dimensions -10, Terminal configuration – Yokogawa EJA530E User Manual

Page 50: Terminal wiring

background image

<9. General Specifications>

9-10

IM 01C25F01-01E

9.4 Dimensions

Model EJ510, EJ530, EJX610A and EJX630A

54

(2.13)

6

(0.24)

5(0.20)

24(0.94)

ø6

(0.24)

172 (6.77)

60(2.36)*

2

12*

3

(0.47)

126(4.96)

165 (6.50)

95(3.74)

Mounting bracket

(optional)

(optional)

Extenal indicator

Conduit connection

Integral indicator

(optional)

188(7.40)

110(4.33)

129(5.08)

12

(0.47)

110(4.33)

39

(1.54)

ш70 (2.76)

ш78(3.07)

Conduit connection

Zero

adjustment

Ground terminal

Open to

atmosphere*

1

Shrouding bolt*

4

2-inch pipe

(O.D. 60.5 mm)

156 (6.14)

With Process connections code 7

With Process connections code 4

With Process connections code 8 and 9

*1: Only for EJ530 and EJX630A whose measurement span code is A, B, or C.

*2: 58 mm(2.28 inch) for measurement span code D.

*3: 11 mm(0.43 inch) for measurement span code D.

*4: Available only when specifying the option code including ATEX, IECEx or TIIS flameproof type.

159(6.26)

Electrical connection

for code 5, 9, A, and D.

F0904.ai

Unit: mm (approx. inch)

SUPPLY

CHECK

or

ALARM

+


+

Terminal Configuration

Power supply and output terminal

External indicator (ammeter) terminal

*5*6

or

Status contact output terminal

*6

(when /AL is specified)
Ground terminal

Communication

terminals (BT200 etc.)

connection hook

Terminal Wiring

*5: When using an external indicator or check meter, the internal

resistance must be 10 Ω or less. A check meter or indicator

cannot be connected when /AL option is specified.

*6: Not available for fieldbus and PROFIBUS communication types.

Check meter

connection hook

*5*6

SUPPLY +

SUPPLY –

CHECK – or ALARM –

CHECK + or

ALARM +

+

F0905.ai