**These parameters are not available if the ramping software option has been ordered (96 _ _ - _ _ _ _ - AA _ _ )
Active if input 2 is present (96
_1 _-_ _ _ _ -_ _ _ _) and Cali-
bration Lock (Lockout Menu)
is not set to [hide].
1608 w
[``no] (0)
[``no] no (0)
[`yes] yes (4)
[b20A] Process Calibra-
tion 2, 20m
A
Store 20.000mA cali-
bration for input 2
process.
Active if input 2 is present (96 _1
_-_ _ _ _ -_ _ _ _) and Calibra-
tion Lock (Lockout Menu) is not
set to [hide].
1608 w
[``no] (0)
[``no] no (0)
[`yes] yes (3)
[b`4A] Process Calibra-
tion 2,
4mA
Store 4.000mA cali-
bration for input 2
process.
Active if input 2 is present (96 _1
_-_ _ _ _ -_ _ _ _) and Calibra-
tion Lock (Lockout Menu) is not
set to [hide].
1608 w
[``no] (0)
[``no] no (0)
[`yes] yes (2)
[b10u] Process Calibra-
tion 2, 10V
Store 10.000V cali-
bration for input 2
process.
Active if input 2 is present (96 _1
_-_ _ _ _ -_ _ _ _) and Calibra-
tion Lock (Lockout Menu) is not
set to [hide].
1608 w
[``no] (0)
[``no] no (0)
[`yes] yes (1)
[b`0u] Process Calibra-
tion 2, 0V
Store 0.000V calibra-
tion for input 2 pro-
cess.
[cin2]
Calibration 2 Menu
**
[Fcty] Factory Page
Active if Calibration Lock (Lock-
out Menu) is not set to [hide].
1603 w
[``no] (0)
[``no] no (0)
[`yes] yes (11)
[A20A] Process Calibra-
tion 1, 20mA
Store 20.000mA cali-
bration for input 1
process.
Active if Calibration Lock (Lock-
out Menu) is not set to [hide].
1603 w
[``no] (0)
[``no] no (0)
[`yes] yes (10)
[A`4A] Process Calibra-
tion 1, 4mA
Store 4.000mA cali-
bration for input 1
process.
Active if Calibration Lock (Lock-
out Menu) is not set to [hide].
1603 w
[``no] (0)
[``no] no (0)
[`yes] yes (9)
[A10u] Process Calibra-
tion 1, 10V
Store 10.000V cali-
bration for input 1
process.
Active if Calibration Lock (Lock-
out Menu) is not set to [hide].
1603 w
[``no] (0)
[``no] no (0)
[`yes] yes (8)
[A`0u] Process Calibra-
tion 1, 0V
Store 0.000V cali-
bration for input 1
process.
Active if Calibration Lock (Lock-
out Menu) is not set to [hide].
1603 w
[``no] (0)
[``no] no (0)
[`yes] yes (7)
[r380] RTD Calibration,
380
Ω
Store 380.00
Ω
cali-
bration for input 1
RTD.