vi

ZOLL M Series CCT

Safety Summary (cont.)

• Do not immerse the M Series CCT device, batteries, cables, or transducers in water, solvents, or

cleaning solutions.

• Do not sterilize M Series unit or accessories except as specifically recommended in ZOLL

manuals. Reusable transducers should be sterilized per the manufacturer’s instructions.

• Connect the ECG-out jack, VGA-out jack, and modem (if available) only to other equipment with

galvanically isolated circuits.

• When using the optional VGA video output connector, test system operation with target video

display device prior to clinical use. Testing should include the Daily Checkout Procedure (refer to

the M Series Operator’s Guide).

• When using the M Series CCT in an aircraft, fluctuations in cabin pressure can affect invasive

blood pressure (IBP) calibration. When cabin pressure fluctuates significantly due to altitude

changes, you must rezero the IBP transducer. We recommend that you rezero the IBP transducer

(1) After takeoff or landing when cabin pressure has stabilized; (2) Whenever the fluctuation in

cabin pressure is equivalent to the change in air pressure that occurs with a 500-foot (152-meter)

change in altitude.

Indications for Use

The M Series CCT indications for use are the same as for the standard M Series. For additional

information, refer to the M Series Operator’s Guide and user inserts for installed options.

Contraindications for Use

The M Series CCT contraindications for use are the same as for the standard M Series. For additional

information, refer to the M Series Operator’s Guide and user inserts for installed options.