Safety summary (cont.), Indications for use, Contraindications for use – ZOLL M Series CCT Defibrillator Rev E User Manual
Page 6

vi
ZOLL M Series CCT
Safety Summary (cont.)
• Do not immerse the M Series CCT device, batteries, cables, or transducers in water, solvents, or
cleaning solutions.
• Do not sterilize M Series unit or accessories except as specifically recommended in ZOLL
manuals. Reusable transducers should be sterilized per the manufacturer’s instructions.
• Connect the ECG-out jack, VGA-out jack, and modem (if available) only to other equipment with
galvanically isolated circuits.
• When using the optional VGA video output connector, test system operation with target video
display device prior to clinical use. Testing should include the Daily Checkout Procedure (refer to
the M Series Operator’s Guide).
• When using the M Series CCT in an aircraft, fluctuations in cabin pressure can affect invasive
blood pressure (IBP) calibration. When cabin pressure fluctuates significantly due to altitude
changes, you must rezero the IBP transducer. We recommend that you rezero the IBP transducer
(1) After takeoff or landing when cabin pressure has stabilized; (2) Whenever the fluctuation in
cabin pressure is equivalent to the change in air pressure that occurs with a 500-foot (152-meter)
change in altitude.
Indications for Use
The M Series CCT indications for use are the same as for the standard M Series. For additional
information, refer to the M Series Operator’s Guide and user inserts for installed options.
Contraindications for Use
The M Series CCT contraindications for use are the same as for the standard M Series. For additional
information, refer to the M Series Operator’s Guide and user inserts for installed options.