Electromagnetic emissions declaration – ZOLL M Series CCT Defibrillator Rev E User Manual
Page 15

ZOLL M Series CCT
9
IEC 60601-1-2 Specifications
This section provides specification tables for the M Series CCT as per IEC 60601-1-2.
Electromagnetic Emissions Declaration
Guidance and manufacturer’s declaration — electromagnetic emissions for the M Series CCT.
The M Series CCT is intended for use in the electromagnetic environment specified below. The customer or
user of the M Series CCT should ensure that it is used in such an environment.
Emissions test
Compliance
Electromagnetic environment – guidance
RF Emissions
CISPR 11
Group 1
The M Series CCT uses RF energy for its internal function only.
Therefore, its RF emissions are very low and are not likely to
cause any interference in nearby electronic equipment.
RF Emissions
CISPR 11
Class B
The M Series CCT is suitable for use in all establishments,
including domestic establishments and those directly connected to
the public low-voltage power supply network that supplies
buildings used for domestic purposes.
Harmonic Emissions
IEC 61000-3-2
Class A
Voltage Fluctuations/
Flicker Emissions
IEC 61000-3-3
Complies