Maintenance, Specifications – ZOLL M Series CCT Defibrillator Rev E User Manual
Page 13

ZOLL M Series CCT
7
Maintenance
Perform all periodic maintenance of the M Series CCT per the M Series Operator’s Guide and the
supplemental inserts for any other options.
Specifications
Unless described below, M Series CCT specifications are identical to M Series specifications. Refer
to the M Series Operator’s Guide for details.
Table: General Specifications
Size: w/ NIBP, IBP, Temp:
w/o NIBP, IBP, Temp:
10.2” (25.9 cm) high x 10.3” (26.2 cm) wide x 8.7” (22.1 cm) deep
8.6” (21.8 cm) high x 10.3” (26.2 cm) wide x 8.7” (22.1 cm) deep
Weight:
17.2 lbs (7.8 kg) with Multi-function cable and battery
19.2 lbs (8.71 kg) with above and paddles
ac Power:
100-120 V, 50/60 Hz; 220-240 V, 50 Hz; 220 VA
dc Input (Optional):
10-29 V, 130 W
Device Classification:
Class I and internally powered per EN 60601-1
Class II and internally powered per EN 60601-1 (DC input only)
Design Standards:
Meets or exceeds UL 2601, AAMI DF-39, AAMI DF-2, IEC 601-2-4,
EN 60601-2-25, and EN 60601-2-27
Patient Safety:
All patient connections are electrically isolated.
Table: Display Specifications
Screen Type:
Active Matrix Color LCD
Screen Size:
6.5” (16.51 cm) diagonal
Number of Pixels:
640 x 480
Sweep Speed:
25 mm/s
Video Output:
Industry standard VGA, 640 x 480, 60 Hz (optional)
Table: Battery Specifications (XL Battery Pack)
Type:
Rechargeable, sealed lead acid
Weight:
3.7 lbs (1.68 kg)
Voltage:
2 volts/cell; 5 cells wired in series
Recharge Time:
7.2 hours or less with integral charger
Operating Time;
For a new, fully charged XL
battery pack at 20°C:
60 defibrillator discharges at maximum energy (200 J), or 2.5 hours
minimum of continuous ECG and SpO
2
monitoring, or 1.5 hours of
continuous ECG, SpO
2
, EtCO
2
, IBP, and Temperature monitoring/pacing at
60 mA, 70 beats per minute.