0 install, 1 open the package, 2 electronic connect 2.2.1 open the terminal – Vestil PM-SCALE User Manual
Page 8: 2 load cell connect, 1+e xc +sen +s ig sh ie ld -s ig -se n -exc

3
2.0 Install
This part will talking about the Installation for IND221 and IND226. Please read this
chapter carefully before install.
2.1 Open the package
Open the package, And check all the parts with the checklist. Make sure no part was
damaged and missing.
Remove the terminal from
its protective package,
.
2.2 Electronic Connect
2.2.1 Open the terminal
The IND221 terminal use 4 screws to lock the front cover.
The front panel of the IND226 terminal is locked in place by four spring clips attached
to the enclosure body. To gain access to the terminal’s PCB for internal wiring and setting
switches, separate the front panel from the enclosure as follows:
Insert the tip of a flat-blade screwdriver into one of the two slots located on the bottom
of the front panel assembly and gently push in toward the enclosure. A “pop” sound is
made when the cover is released.
2.2.2 Load Cell Connect
7 Pins terminal strip.
Pin 1 - +EXC
Pin 2 - +SEN
Pin 3 - +SIG
Pin 4 - Shield
1
+E
XC
+SEN
+S
IG
Sh
ie
ld
-S
IG
-SE
N
-EXC