Changing system settings, Output, Usb output – Teledyne LeCroy WaveStation 3000 Operators Manual User Manual
Page 34: Number format

WaveStation 3000 Function and Arbitrary Waveform Generator
Changing System Settings
Many default WaveStation settings can be modified through the Utility menu. Follow the procedures below to
change Output and System settings.
See the
, and
sections for instructions on using other Utility
menu options.
Also see
Output
This setting changes the output Voltage and Polarity on the generated waveform. For output over 10 V pk-pk,
use the HighZ setting.
1. Press the Utility function button.
2. Press the Output Setup softkey.
3. Press the Load softkey until either 50Ω or HighZ is selected.
4. Press Polarity to select either Normal or Invert.
5. Press Done.
USB Output
The USB Type B port on the back of the WaveStation can be configured to output either:
l
USBRAW
l
USBTMC, used for remote control and WaveStation PC software
1. Press the Utility function button.
2. Press the Interface softkey.
3. Press the USB Setup softkey.
4. Press either USBRAW or USBTMC.
5. Press Done.
Number Format
1. Press the Utility function button.
2. Press the More softkey, then press System and Number Format.
3. Choose a Point style of either period (.) or comma (,). This character is placed after the first thousand.
4. Choose a Separator of:
l
On - places the other Point character (period or comma) between whole units and decimals.
l
Space - uses a space rather than a character to separate whole units and decimals.
l
Off- no separator.
5. Press Done.
28