beautypg.com

Operator’s manual, Table 1. list of required equipment – Teledyne LeCroy AP015 User Manual

Page 9

background image

Operator’s Manual

922254-00 Rev A

5

Table 1. List of Required Equipment

Description

Minimum Requirements

Test Equipment Examples

Wide Band oscilloscope

Minimum 200 MHz bandwidth
ProBus interface equipped
Software Version 7.6.0 or
greater

Teledyne LeCroy LT322

Teledyne LeCroy X-Stream
scope with AP-1M Hi-Z
adapter

Digital Multimeter
(2 required)

DC: 0.1% accuracy 5.5 digit
resolution with Null capability
(referenced measurements)

Agilent Technologies 34401A

Fluke 8842A-09

Keithley 2001

Leveled Sine Wave
Generator

Relative output level accurate
to ±0.3 dB into 50

1 to 100 MHz frequency
range Output adjustable to 18
dBm

Tegam SG503 with TM
series mainframe and
matching output cable.

A semiautomatic software
leveled signal source
calibrated with a power
meter may be substituted.

DC Power Supply

Output adjustable to > 6 V at
1 A

Agilent Technologies
E3610A

Calibration Fixture,
50 Turn Loop

50 Turn loop in series with 0.5
Ω ±0.2% resistor with sense
terminals

Teledyne LeCroy CP015-
CF02

Calibration Fixture,
High Speed Shunt

50

Ω Termination into single

turn loop.

VSWR < 1.25: 1 DC

– 50

MHz

Teledyne LeCroy CP015-
CF01

Calibration Fixture

ProBus Extension Cable

Teledyne LeCroy PROBUS-
CF01

BNC Adapter, T

Male-to-Dual-Female

Pomona 3285

BNC coaxial cable

Male-Male BNC, 50

Ω, 36”

Pomona 5697-36

Banana Plug adapter

BNC Female-to-Banana Plug Pomona 1269

Patch Cables
(4 required)

Male Banana-to-Male
Banana, 12”

Pomona B-12-0 (black)

Pomona B-12-2 (red)