Troubleshooting – SKC Limited Flite 2 User Manual
Page 24
Page 22
901-FTM Issue I
www.skcltd.com
Troubleshooting
Should the Flite 2 pump detect that the battery is at a low
charge state a low battery message will be displayed on the
LCD screen. Charge the battery fully using the correct SKC
charger before use.
BATTERY
TOO LOW
OR
ELAPSED TIME
003:24:17LOW BAT
The Flite 2 pump will display a general fault message on the
LCD screen in the event of any error, as shown.
ELAPSED TIME
000:16:53 FAULT
Possible Fault
Corrective Action
Sample media fault
1. Check media for heavy loading or damage.
2. Rectify problem.
3. Try restarting programming.
Sample train fault
1. Check tubing for blockage or crimping.
2. Rectify problem.
3. Try restarting programming.
Software error (frozen keypad)
1. Disconnect battery for one minute, reconnect.
2. Try restarting programming.
For further assistance contact your supplier or SKC customer care - +44 (0) 1258 480188.