Programming the flite 2 pump – SKC Limited Flite 2 User Manual

Page 14

background image

Page 12

901-FTM Issue I

www.skcltd.com

Programming the Flite 2 Pump

Having set the required pump flow rate disconnect the flow calibrator from the sampling head inlet. Remove
the filter from the sampling head and replace it with a fresh new filter. Fit the sealing cap to the sampling head
inlet to prevent contamination of the filter prior to sampling.

3) Manual Operation

Having set the pump flow rate to the desired level, the ‘RUN
PARAMETER’ menu screen is displayed.

SET FLOW USING
>>>>>>>

Æ

RUN PARAMETER

MANUAL/time/exit

For simple non-timed, manual operation select the
‘MANUAL’ option, and press

to display the ‘REVIEW

DATA ?’ menu screen. Refer to section - 5) Reviewing
Sampling Programs, for further details. Alternatively, select
the ‘EXIT’ option, and press

to cancel the sample

program and return to the ‘SET UP NOW ?’ menu screen.

RUN PARAMETER

MANUAL/time/exit

Æ

REVIEW DATA ?

yes/NO

4) Setting a Timed Run

Having set the pump flow rate to the desired level, from the
‘RUN PARAMETER’ menu screen select the ‘TIME’ option,
and press

to display the next menu screen.

RUN PARAMETER

manual/TIME/exit

Æ

ENTER RUN TIME ?

YES/no