beautypg.com

SKC Limited Flite 2 User Manual

Page 13

background image

www.skcltd.com

901-FTM Issue I

Page 11

Programming the Flite 2 Pump

Select the ‘YES’ option in the ‘SET UP NOW ?’ menu
screen, and press

to display the next menu screen.

SET UP NOW ?

YES/no

Æ

SET FLOW NOW ?

YES/no

Ensure that the ‘YES’ option is selected and press

to

display the next menu screen.

Æ

SET FLOW USING
>>>

Press

to increase the pump flow rate. More arrow

symbols are displayed on the LCD screen to indicate the
increased flow rate.

Æ

SET FLOW USING
>>>>>>>>>>>>

Press

to decrease the pump flow rate. Less arrow

symbols are displayed on the LCD screen to indicate the
reduced flow rate.

Æ

SET FLOW USING
>>>>>>>

Observing the reading of the flow calibrator, use the

and

keys to set the required flow rate. When the pump is

running at the required flow rate press

to save the flow

rate setting and display the next menu screen.
If the flow rate is not entered within 60 seconds, the following
screens are displayed and the flow rate must be set again.

NO FLOW ENTERED

TRY AGAIN

Æ

SET FLOW NOW ?

YES/no