Skc pump history window history display – SKC DataTrac for the Pocket Pump User Manual
Page 22
20
SKC Pump History Window
History Display
The SKC Pump History display (Figure 29) shows the record or history
of all operations performed by the pump.
A history will remain on screen and in the pump memory until it is
cleared. A history includes the following data:
Readout
State of the Pump
Flow .......................pump in constant fl ow
CP ..........................pump in constant back pressure
Hold ......................pump in hold
Flow Fault.............fl ow fault occurred while in constant fl ow
CP Flow Fault ......fl ow fault occurred while in constant back pressure
Fault ......................fault of unknown origin
Low Batt ery ..........batt ery depleted
Hibernate...............a low power state that activates aft er the pump has been in hold for 24 hours. See explanation
of hibernate on page 25.
Rate ........................fl ow rate in mL/min or the back pressure in inches H
2
O or mm Hg
Start Date ..............starting date of the program
Start Time .............starting time of the program
Volume (Liters) ....fl ow rate multiplied by the duration
Accum Volume ....sum of all previous volumes (liters)
Duration ...............total run time of the program
A discrepancy may occur between the pump display and the history during fl ow fault. After a fl ow fault,
the pump will enter Hold mode and will attempt to restart every 5 minutes. If the fault has not been
corrected within 15 seconds, the pump will return to Hold mode. The volume of air pumped during
attempted restarts will appear on the pump LCD, which updates every second, but will not appear in
the history.
Figure 29. History Display