Inserting a Collection Filter into the IMPACT Sampler

The IMPACT Sampler will arrive already assembled. Disassemble it to insert
collection fi lter.

Unscrew the inlet from the exhaust.

Remove the fi lter cassett e.

3a. Use the fi lter cassett e opener

to open the fi lter cassett e.


3b. Slide the fi lter cassett e horizon-

tally into the “U” of the opener
until the two halves of the cas-
sett e loosen. Gently pull halves
apart.

Ensure the stainless steel support

screen is in place in the bott om of the

fi lter cassett e.

Using forceps, place a preweighed

47-mm fi lter on the support screen.

Press the fi lter cassett e top into the

fi lter cassett e bott om.

Reinsert the cassett e into the impactor.