SKC 100-3901 DPS (Deployable Particulate Sampler) System User Manual
Page 8
![background image](/manuals/349570/8/background.png)
5
Inserting a Collection Filter into the IMPACT Sampler
The IMPACT Sampler will arrive already assembled. Disassemble it to insert
collection fi lter.
Unscrew the inlet from the exhaust.
Remove the fi lter cassett e.
3a. Use the fi lter cassett e opener
to open the fi lter cassett e.
3b. Slide the fi lter cassett e horizon-
tally into the “U” of the opener
until the two halves of the cas-
sett e loosen. Gently pull halves
apart.
Ensure the stainless steel support
screen is in place in the bott om of the
fi lter cassett e.
Using forceps, place a preweighed
47-mm fi lter on the support screen.
Press the fi lter cassett e top into the
fi lter cassett e bott om.
Reinsert the cassett e into the impactor.
1
7
Technical Tidbit
•
Use forceps to carefully
insert and remove the
collection filter. See
Accessories for forceps.