10
Leland Legacy Quick Guide
Terms »
Star button
• Scrolls through run time data and Setup options
Up and down arrow buttons
• Toggle between display choices and increase or decrease sampling parameters in Setup
Button sequence
= press buttons individually
[] = press simultaneously
= security code, always press in sequence
Security code
• Prevents unauthorized changes to the pump’s sampling program
Programming Sequences »
• To activate pump (e.g., to change pump from Sleep to Hold):
Press any button.
• To change pump from Hold to Run or Run to Hold:
Press [].
• To reset accumulated data:
Press [], then . Press until CLr displays then press []; press until End displays
then press [].
• To set pump fl ow rate:
Press [], then . Flow rate and SET fl ash. Press or to change fl ow rate. Press until
End appears then press [] to save setting and place pump in Hold.
• To calibrate fl ow rate with standard calibrator:
Press [], then . Flow rate and SET fl ash. Press or to change fl ow rate. Press once.
ADJ displays. Press or until desired fl ow rate is indicated on calibrator. When fi nished, press until
End displays then press [] to save new setting and place pump in Hold. For CalChek Calibration, see
operating instructions.
• To change temperature scale from F to C or C to F:
Press [], then . Press until temperature displays. Press or to switch units; press
until End displays then press [] to save new setting.
• To change atmospheric pressure scale (mm, mb, In):
Press [], then . Press until pressure displays then press or to switch units; press
until End displays then press [] to save new setting.
• To change time scale (12 Hr/24 Hr/Dela):
Press [], then . Press until 12 Hr, 24 Hr, or Dela displays then press or to switch
units; press until End displays then press [] to save new setting. To set delayed start (Dela), see
operating instructions.
• To change clock:
Press [], then . Press until clock displays then press or to change fl ashing hour; press
to move to minutes and or to change setting. Press until End displays then press [] to save
new setting.
• To change the sampling time function:
Press [], then . Press until ST L/min displays then press to change fl ashing digit; press
until End displays then press [] to save new setting. To delete, follow above steps and press until
0 appears. Exit Setup.
Note: When in Setup, choosing Esc instead of End will exit Setup without saving new settings.