beautypg.com

Leland legacy quick guide, Terms, Programming sequences – SKC 100-3901 DPS (Deployable Particulate Sampler) System User Manual

Page 13

background image

10

Leland Legacy Quick Guide

Terms »

Star button

• Scrolls through run time data and Setup options

Up and down arrow buttons 

• Toggle between display choices and increase or decrease sampling parameters in Setup

Button sequence

 = press buttons individually

[] = press simultaneously



= security code, always press in sequence

Security code 

• Prevents unauthorized changes to the pump’s sampling program

Programming Sequences »

• To activate pump (e.g., to change pump from Sleep to Hold):

Press any button.

• To change pump from Hold to Run or Run to Hold:

Press [].

• To reset accumulated data:

Press [], then . Press  until CLr displays then press []; press  until End displays

then press [].

• To set pump fl ow rate:

Press [], then . Flow rate and SET fl ash. Press  or  to change fl ow rate. Press  until

End appears then press [] to save setting and place pump in Hold.

• To calibrate fl ow rate with standard calibrator:

Press [], then . Flow rate and SET fl ash. Press  or  to change fl ow rate. Press  once.

ADJ displays. Press  or  until desired fl ow rate is indicated on calibrator. When fi nished, press  until

End displays then press [] to save new setting and place pump in Hold. For CalChek Calibration, see

operating instructions.

• To change temperature scale from F to C or C to F:

Press [], then . Press  until temperature displays. Press  or  to switch units; press 

until End displays then press [] to save new setting.

• To change atmospheric pressure scale (mm, mb, In):

Press [], then . Press  until pressure displays then press  or  to switch units; press 

until End displays then press [] to save new setting.

• To change time scale (12 Hr/24 Hr/Dela):

Press [], then . Press  until 12 Hr, 24 Hr, or Dela displays then press  or  to switch

units; press  until End displays then press [] to save new setting. To set delayed start (Dela), see

operating instructions.

• To change clock:

Press [], then . Press  until clock displays then press  or  to change fl ashing hour; press

to move to minutes and  or  to change setting. Press  until End displays then press [] to save

new setting.

• To change the sampling time function:

Press [], then . Press  until ST L/min displays then press  to change fl ashing digit; press

until End displays then press [] to save new setting. To delete, follow above steps and press  until

0 appears. Exit Setup.

Note: When in Setup, choosing Esc instead of End will exit Setup without saving new settings.

SKC Inc., 863 Valley View Road, Eighty Four, PA 15330 • www.skcinc.com