MEDIA PREPARATION

Collection Filters: Equilibrate and pre-weigh fi lters in a clean environment
according to appropriate procedures. Record the weight as the pre-sample
weight.

Impaction Disc: Ready-to-use pre-oiled disposable plastic impaction discs are included
with the system (qty. 25). Replacement pre-oiled plastic discs are available as SKC
Cat. Nos. 225-395 (qty. 25) and 225-395A (qty. 50).
Using an oiled impaction disc
reduces particle bounce.

37-mm fi lters (quartz or PTFE) may be used if chemical analysis of larger par-
ticles is desired.

IMPACTOR PREPARATION

Cleaning

For optimum performance, the IMPACT Sampler inlet, exhaust, and fi lter cas-
sett e should be cleaned aft er fi ve runs or upon a noticeable buildup of material.
This will remove oil buildup from the top of the fi lter cassett e and other residue
built up from frequent sampling. Disassemble the impactor and wash parts in
water with a liquid detergent or soap. Rinse and air dry all parts thoroughly in a
clean environment.

Do not place any mechanical object in the inlet nozzles.

O-ring Care

Visually inspect the condition of the BUNA-N exhaust O-ring (see illustration on
page 3 for location
). Ensure the O-ring surface is smooth (i.e., without cracks, cuts,
or other damage). Ensure the O-ring is fi tt ed properly in its channel. Replace the
exhaust O-ring if there is apparent damage, stretching, or thinning. It is recom-
mended that the PTFE inlet O-ring be replaced by the manufacturer only.