Sampling – SKC 100-3901 DPS (Deployable Particulate Sampler) System User Manual

Page 15

background image

12

Sampling

1. Replace the representative sample media used for calibration with new, pre-

weighed media (see Media Preparation and Impactor Preparation).

2. Mount the bracket at the desired location and at breath-

ing zone height (6 feet or 2 meters) using wire ties or
other fasteners. Mount the impactor on the mounting
bracket by threading the clamp knob into the bott om of
the impactor.

3. Insert the screw on the rain cover into the screw hole in

top of the impactor inlet and rotate the cover until tight.

4. Turn on the pump and record sample start time, ambient

temperature, ambient pressure, and other pertinent data.

Sample start time and duration can be programmed
into the Leland Legacy Sample Pump in advance and
sampling may be started manually or automatically.

Record all necessary data before disconnecting the
pump from the batt ery and reconnecting to the new
batt ery.

5. Aft er desired sample time has elapsed, turn off the pump and record the

sample stop time, total volume, ambient temperature, ambient pressure, and
other pertinent data.

6. Remove the rain cover, reinstate the calibration

train and media, and verify the fl ow rate (see Cali-
bration
).

7. Reach inside the case and press quick-connect

release while pulling the tubing from the inlet.
Remove tubing from the impactor.

Technical Tidbits

The supplied rain cover should be used for all outdoor sampling.

Keep the Leland Legacy sample pump inside the Pelican case and the
case closed during sampling to protect the sample pump from weather.

3

4

7

Quick-connect

release inside

case

Tubing with

quick-connect

fi tting