11

CALIBRATION AND SAMPLING

Calibration

Calibrate pump fl ow rate with the IMPACT Sampler in line (loaded with
representative fi lter and impaction disc). See pump and calibrator operating
instructions.

Set the pump fl ow rate to 10 L/min (see Leland Legacy
Quick Guide on page 10). Ensure the pump has run for
5 minutes before proceeding with calibration. Ensure the
rain cover is removed from the inlet and that the impactor
is loaded with representative media and fully assembled (see
Inserting a Collection Filter into the IMPACT Sampler and
Inserting an Impaction Disc into the IMPACT Sampler).

Screw the calibration adapter onto

the impactor inlet.

Calibrate pump fl ow rate with a calibrator. Adjust the

fl ow rate until the calibrator displays 10 L/min (see Leland

Legacy Quick Guide on page 10). Record the fl ow rate.

Reset the accumulated data before sampling. See pump and

calibrator operating instructions.

1

When calibration is completed,

disconnect the calibrator and tubing

from the calibration adapter. Remove

the calibration adapter from the

impactor. Remove the representative

media and replace with fresh media

for sampling.

Use a short length of calibration tubing to connect the inlet of the

calibration adapter to the outlet of a calibrator to form a calibration train.

Unscrew the quick-connect plug. Use tubing with a quick-connect fi tt ing to

att ach the case (pump) inlet to the outlet of impactor.

Ensure O-ring is installed on the quick-connect fi tt ing before
inserting it into the inlet. Absence of the O-ring can aff ect
measurements.

Inlet

Quick-

connect

plug

Inlet line

to pump

Quick-

connect

fi tting on

tubing

Inlet to

pump

IMPACT outlet

Calibrator

outlet

Calibration

adapter