Oxygen Analyzer

Maintenance

Teledyne Analytical Instruments

59

Figure 5.1 Adding Water into the BDS sensor

WARNING:

THE SENSOR USED IN THE MODEL BDS 3960
OXYGEN ANALYZER USES ELECTROLYTE WHICH
CONTAINS POTASSIUM HYDROXIDE, THAT CAN BE
HARMFUL IF TOUCHED, SWALLOWED, OR
INHALED. AVOID CONTACT WITH ANY FLUID OR
POWDER IN OR AROUND THE UNIT. WHAT MAY
APPEAR TO BE PLAIN WATER COULD BE THE
ELECTROLYTE. IN CASE OF EYE CONTACT,
IMMEDIATELY FLUSH EYES WITH WATER FOR AT
LEAST 15 MINUTES. CALL PHYSICIAN. (SEE
APPENDIX, MATERIAL SAFETY DATA SHEET.)

5.3

Fuse Replacement

1. Place small screwdriver in notch, and pry cover off, as shown in

Figure 5-2. Remove fuse holder.

2. Replace fuses. Use 2A 250 VAC 5x20 mm slow-blow. Install

one at each side of fuse holder