Oxygen Analyzer

Installation

Teledyne Analytical Instruments

35

4–20 mA DC % Range:

Current increases linearly with
increasing oxygen, from 4 mA at 0
ppm to 20 mA at full scale ppm. (Full
scale = 100% of programmable range.)

4–20 mA dc Range ID:

8 mA = Low Range, 12 mA = Medium
Range, 16 mA = High Range, 20 mA
= 100ppm.

Table 3-1: Analog Output Connections

Pin

Function

3

+ Range ID, 4-20 mA, floating

4

– Range ID, 4-20 mA, floating

5

+ % Range, 4-20 mA, floating

6

– % Range, 4-20 mA, floating

8

+ Range ID, 0-1 V dc

23

– Range ID, 0-1 V dc, negative ground

24

+ % Range, 0-1 V dc

7

– % Range, 0-1 V dc, negative ground

Alarm Relays: The nine alarm-circuit connector pins connect to the
internal alarm relay contacts. Each set of three pins provides one set of
Form C relay contacts. Each relay has both normally open and normally
closed contact connections. The contact connections are shown in Table
3-2. They are capable of switching up to 3 amperes at 250 VAC into a
resistive load. The connectors are:

Threshold Alarm 1:

Can be configured as high (actuates when concentration
is above threshold), or low (actuates when concentration
is below threshold).

Can be configured as failsafe or nonfailsafe.

Can be configured as latching or nonlatching.

Can be configured out (defeated).