Introduction

BDS 3960

Teledyne Analytical Instruments

16

Access Drawer: For access t o the BDS S ensor And pr essur e regul ator.
T he f ront panel sli des out when the four thumb screws ar e loosened.
Openi ng the int er ior gives access to most ci rcui t boards t oo.

1.5

Rear Panel (Equipment Interface)

The rear panel, shown in Figure 1-2, contains the gas and electrical

connectors for external inlets and outlets. The connectors are described
briefly here and in detail in Chapter 3 Installation. Except for gas, AC
power, and RS-232 communications I/O, all user connections for analog
output, alarms, calibration and remote probe contacts are available
through the 50 pin equipment interface connector. User connections
made at the rear panel include:

Power Connection

110 VAC power source (220 VAC
optional.

Gas Inlet and Outlet

One inlet and one exhaust out.

Moisture By-pass

Additional vent line for condensate

Analog Outputs

0–1 VDC oxygen concentr ation plus
0-1 VDC r ange I D, and isolated
4–20 mA DC oxygen concentr ation
plus 4-20 mA DC range ID.

Alarm Connections

2 concentration alarms and 1
system alarm.

Remote Probe

Used in the BDS 3960 for
controlling external solenoid
valves only.

Remote Span/Zero

Digital inputs allow external
control of analyzer calibration.

Calibration Contact

To notify external equipment that
instrument is being calibrated and
readings are not monitoring
sample.•

Range ID Contacts

Four separate, dedicated, range
relay contacts. Low, Medium,
High, Cal.