Installation

BDS 3960

Teledyne Analytical Instruments

30

Installation

Installation of the Model BDS 3960 Analyzer includes:

Unpacking

Mounting

Gas connections

Electrical connections

Filling the Sensor with Electrolyte.

Testing the system.

3.1

Unpacking the Analyzer

Although the analyzer is shipped complete, certain parts, such as

the electrolyte, are wrapped separately to be installed on site as part of
the installation. Carefully unpack the analyzer and inspect it for damage.
Immediately report any damage or shortages to the shipping agent.

3.2

Mounting the Analyzer

The Model BDS 3960 is for indoor use in a general purpose area. It

is NOT for hazardous environments of any type.

T he standard model is designed for f l ush panel m ounti ng. F igur e 3-1

i s an i l lust r at ion of the BDS 3960 st andard fr ont panel and mount ing
bezel . There ar e four mount ing hol es—one in each corner of t he ri gi d
f rame. The dr awings sect i on i n the r ear of thi s manual contai ns out l ine
dim ensi ons and mount ing hol e spaci ng di agram s.