Introduction

BDS 3960

Teledyne Analytical Instruments

14

Three user definable output ranges (from 0-100 ppb through
0-100 ppm) allow best match to users process and
equipment, plus a fixed 100 ppm over range.

Auto Ranging allows analyzer to automatically select the
proper preset range for a given measurement. Manual
override allows the user to lock on to a specific range of
interest.

Two adjustable concentration alarms and a system failure
alarm.

Two way RFI protection.

Four analog outputs: two for measurement (0–1 VDC and
Isolated 4–20 mA DC) and two for range identification.

Convenient and versatile, steel, flush-panel or rack-
mountable case with slide-out electronics drawer.

1.4

Front Panel (Operator Interface)

The standard BDS 3960 is housed in a rugged metal case with all

controls and displays accessible from the front panel. See Figure 1-1.
The front panel has the pocket PC, a digital meter, an alphanumeric
display, and a window for viewing the sample flowmeter.

There are no keys to press on the front panel. All interface with

the analyzer is done through the touchscreen of the pocket PC. The main
functions on the touchscreen are listed below.

System

Perform system-related tasks
(described in detail in chapter 4,
Operation.).

Span

Span calibrate the analyzer.

Zero

Zero calibrate the analyzer.

Alarms

Set the alarm setpoints and
attributes for Alarm 1 and Alarm 2.

Range

Set up the 3 user definable ranges
for the instrument.

Quit

Quit analyzer application.