Teledyne BDS-3960 - Oxygen analyzer User Manual
Page 48

Maintenance
BDS 3960
Teledyne Analytical Instruments
48
3. After the sensor has been purged for at least two weeks and the
baseline is stable, monitor the oxygen reading and ambient
temperature over a minimum period of 24 hours.
Take the maximum and minimum oxygen readings, and the
maximum and minimum temperature readings.
4. Calculate the coefficient using the relation:
Coefficient = (O
2
max
- O
2
min
) ÷ (Temp
max
- Temp
min
)
For example:
In a 24 hour run:
O
2
max
= 3.55 ppb
O
2
min
= 1.75 ppb
Temp
max
= 24.5 degrees C.
Temp
min
= 22.1 degrees C.
Coefficient = (3.55-1.75)ppb ÷ (24.5-22.1) °C = 0.75 ppb/°C.
This value is currently set to zero, since the sensor is
temperature controlled and should be left at zero unless qualified
personnel give instructions to change it.
4.3.2.4 S
ET
D
EFAULTS
The Set Defaults button will reset all parameters in the software,
such as calibration, alarm and range settings, Filter, Gas Factor,
temperature coefficient, etc. to their factory default values. Some of the
default values are listed below:
Table 4-2: Parameter Default Values
Parameter
Default Value
Zero Offset
0
Span Gas
8.00 PPM
SpanFactor
24.42 (Default span factor =
100000 ppb/ 4095 ADC counts)
- 1220 - Multipoint flammable gas and vapor detection system (50 pages)
- 212R - Thermal conductivity analyzer (28 pages)
- 235 - Thermal conductivity analyzer (38 pages)
- 275R - Portable turbine generator purge gas analyzer (21 pages)
- 2000A-EU - General purpose thermal conductivity analyzer (86 pages)
- 2000XTC - Thermal conductivity analyzer (40 pages)
- 2010A - Split architecture thermal conductivity analyzer (110 pages)
- 2010B - Split architecture thermal conductivity analyzer (98 pages)
- 2020 - Explosion proof thermal conductivity analyzer (80 pages)
- 2120 - Trace Nitrogen in Argon Analyzer (66 pages)
- 2120XL - Trace Nitrogen Analyzer (85 pages)
- 2230R - Process Hydrogen Analyzer (26 pages)
- 2240 – Portable Handheld Hydrogen Leak Detector, 3rd generation (updated 1/31/11) (30 pages)
- 2240 - Portable Handheld Hydrogen Leak Detector, 3rd generation (revision 2/29/08) (40 pages)
- 2240 – Portable Handheld Hydrogen Leak Detector, 2nd generation (13 pages)
- 2750 - Portable turbine generator gas analzyer (40 pages)
- 300P - Percent oxygen analyzer (24 pages)
- 306WA - Analog trace oxygen analyzer (46 pages)
- 311 - Portable trace oxygen analyzer (19 pages)
- 311D - Portable trace oxygen analyzer with digital meter (18 pages)
- 311XL - Portable trace oxygen analyzer (18 pages)
- 316RA / RB / RAD / RBD - Oxygen analyzers (24 pages)
- 319R - Oxygen analyzer (23 pages)
- 320 Series - Portable oxygen detectors (24 pages)
- 326, 327 and 328 - Oxygen analyzers (45 pages)
- 329R - Oxygen analyzer (22 pages)
- 335 - Analog control room monitor for personnel safety (24 pages)
- 356WA - Analog trace oxygen analyzer (42 pages)
- 3000MA - Paramagnetic oxygen analyzer (63 pages)
- 3000MA - Paramagnetic oxygen analyzer Addendum (2 pages)
- 3000MB - Paramagnetic oxygen analyzer (59 pages)
- 3000PA - General purpose percent oxygen analyzer (69 pages)
- 3000PAEU - General purpose percent oxygen analyzer (78 pages)
- 3000PB - Bulkhead mount percent oxygen analyzer (82 pages)
- 3000TA - General purpose trace oxygen analyzer (75 pages)
- 3000TA-EU - General purpose trace oxygen analyzer (89 pages)
- 3000TA-XLEU - Trace oxygen analyzer (108 pages)
- 3000TB - Bulkhead mount trace oxygen analyzer (78 pages)
- 3000TB-XL - Trace oxygen analyzer (78 pages)
- 3000ZA - Trace oxygen analyzer (81 pages)
- 3000ZA-3X - Trace oxygen analyzer (72 pages)
- 3000ZA2G - Zirconium oxide analyzer (72 pages)
- 3000 Ultra Trace - PPB oxygen analyzer (72 pages)
- 3010MA - Paramagnetic oxygen analyzer, includes 0-100% range (88 pages)
- 3010MA – Paramagnetic oxygen analyzer, no 0-100% range – (superceded) (88 pages)