Oxygen Analyzer

Maintenance

Teledyne Analytical Instruments

55

as 0–1000 ppb, range 3 cannot be set as 0– 500 ppb since
it is lower than range 2.

Note: Refer to Section 4.3.2.2 to find maximum working range.

4.6.2

Fixed Range Analysis

The autoranging mode of the instrument can be overridden, forcing

the analyzer DC outputs to stay in a single predetermined range.

To switch from autoranging to fixed range analysis, click on the list

box of the main screen, as shown on figure 4.3, then select on the range
of interest.

4.8

Signal Output

The standard Model BDS 3960 Oxygen Analyzer is equipped with

two 0–1 VDC analog output terminals (one concentration and one range
ID), and two isolated 4–20 mA DC current outputs (one concentration
and one range ID) accessible from the 50-pin equipment interface
connector located on the back panel.

See Rear Panel in Chapter 3, Installation, for illustration and pin

configuration of the interface connector.

The signal output for concentration is linear over the currently

selected analysis range. For example, if the analyzer is set on a range
that was defined as 0–100 ppb O

2

, then the output would be as shown in

Table 4-2.

Table 4-2: Linear Output for a 0-100 ppm O2 Range

Voltage Signal

Current Signal

ppb O

2

Output (VDC)

Output (mA DC)

0

0.0

4.0

10

0.1

5.6

20

0.2

7.2

30

0.3

8.8

40

0.4

10.4

50

0.5

12.0