Teledyne 4030 - Total Hydrocarbon gas analyzer User Manual

Page 7

background image

Series 4030

Teledyne Analytical Instruments v

Procedure for Removal of Internal Inaccessible
Shock Hazards

CAUTION:

SERVICING OR MAINTENANCE OF THE ANALYZER
SHOULD ONLY BE DONE BY SUITABLE TRAINED
PERSONNEL. TO AVOID THESE INACCESSIBLE
HAZARDOUS VOLTAGES WHEN SERVICING THE
SERIES 4060, PERFORM EACH OF THE FOLLOWING
STEPS, IN THE ORDER GIVEN, BEFORE SERVICING
BEGINS:

Switch off the power to the Analyzer and remove the main (line)

power cord from the Analyzer.

Remove all external voltages from the connections to the alarm

contacts.

Wait one minute.
Discharge the anode supply voltage.

a. Connect one end of an insulated (to 1000 VDC or more)

clip lead to Analyzer chassis ground (the standoff for the
upper right corner of the mother PCB).

b. Put one end of a 500V rated 1000 ohm resistor in the

other end of the clip lead.

c. Check the voltage between chassis ground (the standoff

for the upper right corner of the mother PCB) and the top
side of R2 at PCB number B74671. It should be between
-5VDC and +5VDC. If is in that range, the inaccessible
hazardous voltage removal procedure is completed, if not
repeat steps 4.a and 4.b.

If it is absolutely necessary to work inside the instrument with power

on, use the ONE HAND RULE:

Work with one hand only.
Keep the other hand free without contacting any other object. This

reduces the possibility of a ground path through the body in case of
accidental contact with hazardous voltages.

WARNING:

THIS INSTRUMENT IS DESIGNED TO BE OPERATED
IN A NONHAZARDOUS AREA. THE ANALYZER USES
HYDROGEN GAS AND/OR OTHER COMBUSTIBLE
GASES IN ITS OPERATION. THIS EQUIPMENT, IF
NOT USED AND MAINTAINED PROPERLY CAN BE
AN EXPLOSION HAZARD. THE ANALYZER,