Teledyne Analytical Instruments 26

CAUTION

: THE GASES USED MUST BE OF THE

HIGHEST QUALITY, ULTRA ZERO GRADES, AS SHOWN
BELOW. FAILURE TO DO SO WILL RESULT IN
CONTAMINATION AND FAILURE TO DETECT LOW
CONCENTRATIONS OF HYDROCARBONS.

AIR: USE ULTRA ZERO GRADE AIR WITH THC LESS THAN
0.1 PPM. DO NOT USE OIL PUMPED AIR UNDER ANY
CIRCUMSTANCES.

FUEL: USE ULTRA HIGH PURITY (UHP) GRADE HYDROGEN
WITH THC LESS THAN 0.5 PPM.

CARRIER GAS: USE ULTRA HIGH PURITY GRADE
NITROGEN GAS, WITH THC LESS THAN 0.1 PPM.

3.3.5.2

E

FFLUENT

All the gases introduced into the detection cell vent from one fitting

at the rear of the analyzer. TAI recommends that the cell be permitted to
vent directly to the atmosphere wherever possible.

If a vent line is required, the installation must include a drop-out pot

to collect the water that is formed by the burning of the hydrogen fuel.
The vent line must be constructed so that water and dirt cannot collect in
it.

3.3.5.3

S

AMPLE

B

YPASS

V

ENT

The sample bypassed by the back-pressure regulation system vents

from a separate port at the rear of the analyzer. If a vent line is required,
it must be installed so that water and dirt cannot accumulate in it.

3.3.5.4

F

UEL AND

A

IR

C

ONNECTIONS

The fuel used to provide combustion should be Hydrogen gas, zero

quality with certified THC of less than 0.5 ppm and a pressure of 30
psig. The compressed air, ultra zero gas quality with THC less than 0.1
ppm should be used at a controlled pressure of 30 psig Connect the fuel
and air sources to the instrument according to the gas connection
diagram included at the back of this manual.