Teledyne 4030 - Total Hydrocarbon gas analyzer User Manual
Page 31
Operation
Teledyne Analytical Instruments 29
the temperature controller is maintaining the temperature set point) and
the red flame failure lamp is on. See Figure 4-1.
The Series 4030 will automatically attempt a flame ignition
sequence following the warm up period, which has been preset at the
factory. If the ignition process fails, the instrument will attempt to ignite
the flame 4 more times. If it continues to fail after the fifth attempt, a
flame failure message will appear on the display. If this occurs refer to
Section 4.3.1.
4.3.1 Verification of the Flame Guard Circuit
The operation of the flame guard circuit has been checked at the
factory, but should be re-verified during start-up. Use the following
procedure after ignition of the flame has been achieved:
1. Turn off the fuel at the supply cylinder.
2. Observe the fuel pressure gauge on the analyzer control panel. The
gauge indication will decay as the fuel in the line is exhausted.
When the gauge reading reaches the vicinity zero, the flame will
be extinguish as the fuel solenoid shuts off the fuel supply. The
analyzer will automatically try to re-ignite. After 5 attempts, it will
display: flame failure, check air, fuel and the flame failure LED
will be on.
3. Open the cylinder supply valve and re-ignite the flame by
pushing the UP and DOWN buttons simultaneously.
4.3.2 Ignition and/or Flame Guard Circuit Failure
If the flame ignition or guard circuits do not operate as described in
the above two sections, set the instrument fuel regulator to the
recommended pressure.
4.3.3 Sample Pump
A built-in Sample Pump is provided to draw Sample Gas from the
Sample line. The gas flow rate should be maintained at about 1000 to
1200 cc/minute using the Front Panel FLOW CONTROL valve
(clockwise increases flow).
4.4 Analyzer
Operation
Although the Model 4030 has been programmed for your application
at the factory, it can be further configured at the operator level.