Operation

Teledyne Analytical Instruments 27

Operation

4.1 Placing the System in Operation

1. Plug the power in the AC power inlet in the back of the Analyzer

and the Sample Selector Unit.

2. Allow at least 2 hours warm-up (heat up sensor & sample

system) after making the air adjustment described below. Warm
up time is set by the software at the factory but may be
overridden by pressing the key.

DO NOT attempt to ignite the flame during warm up
countdown
. Condensation will occur.

4.2 Activating the Support Gases

4.2.1 Air

1. Set the air tank regulator to 30 psig.
2. Adjust the Analyzer Air regulator until the air pressure gauge

reads the recommended air pressure of 9 psig.

After the air is flowing through the sensor and warm-up time has

been completed, activate the following gases:

4.2.2 Carrier Gas

Set the Carrier gas tank regulator to 50 psig and adjust the Analyzer
Sample regulator until its (sample) pressure gauge reads the
recommended sample pressure of 9.5 psig.

4.2.3 Span Gas

The Span gas containing certified mixture of methane in

air, recommended range 6 to 8 ppm methane in zero grade
air, may be connected to the Span Manifold or the ¼ inch
connection on the back of the Sample Selector Module.