Teledyne 4030 - Total Hydrocarbon gas analyzer User Manual

Page 24

background image

Installation

Teledyne Analytical Instruments 22

Table 3-2: Alarm Relay Contact Pins

Pin Contact

45

Threshold Alarm 1, normally closed contact

28

Threshold Alarm 1, moving contact

46

Threshold Alarm 1, normally open contact

42

Threshold Alarm 2, normally closed contact

44

Threshold Alarm 2, moving contact

43

Threshold Alarm 2, normally open contact

36

System Alarm, normally closed contact

20

System Alarm, moving contact

37

System Alarm, normally open contact





3.3.3.6

R

ANGE

ID

R

ELAYS

There are three dedicated Range ID relay contacts. They are

assigned to relays in ascending order—Low range is assigned to Range
1 ID, Medium range is assigned to Range 2 ID, and High range is
assigned to Range 3 ID. Table 3-6 lists the pin connections. Contacts
are normally open, and they close when 4040 switches to that particular
range.

Table 3-3: Range ID Relay Connections

Pin

Function

21

Range 1 ID Contact

38

Range 1 ID Contact

22

Range 2 ID Contact

39

Range 2 ID Contact

19

Range 3 ID Contact

18

Range 3 ID Contact

34

Not

Used

35

Not

Used