3 flame ignition – Teledyne 4030 - Total Hydrocarbon gas analyzer User Manual

Page 30

background image

Operation

Teledyne Analytical Instruments 28



4.2.4 Fuel

1. Open the main valve on the tank and set the fuel tank
pressure regulator to 30 psig.
3. Adjust the Fuel regulator until its pressure gauge reads the

recommended pressure of 3.5 psig.

Note: Adjust fuel settings only when the red LED (flame failure

light) is off.




TELEDYNE ANALYTICAL INSTRUMENTS

BYPASS FLOW

SPAN GAS

TEMP. GAUGE

SAMPLE

SPAN

Figure 4-1: Front Panel View of Regulator and Gages

4.3 Flame

Ignition

Observe that after warm up count down timer reaches zero (timer to

preheat the sensor), the amber heater lamp is blinking (indicating that