Appendix

3000ZA

Teledyne Analytical Instruments

To improve the system response, a by-pass can be added to increase

the sample flow rate to the analyzer by a factor of ten. A by-pass provides
a sample flow path around the analyzer of 2 - 18 SCFH. typically.

CALIBRATION GAS:

3000 series analyzer requirements for units with Auto-Cal options

are as follows:

The customer must supply control valves (or restrictors) for any SPAN

or ZERO gas source which is attached to the Auto-Cal ports. The valve
should be adjusted to the same flow rate as the sample gas. When restrictors
are used, the gas pressure must be adjusted to achieve the proper flow rate.

OPERATION WITHOUT A RESTRICTOR DEVICE:

Operation without a restrictor device is not recommend as

mentioned above. A 3000TA without any flow restrictor device was
tested on 11-19-97. This results in a flow rate of 2.4 SLPM @ 1 PSIG.
This is a cv of 0.023 for the standard sample system.

REFERENCE: FLOW_1.XLS & FLOW_2.XLS for information on
flow rates at various pressures.

TAI PART NUMBERS

Restrictor kit:

A68729

Union (SS)

U11

LP Restrictor

R2323 (low pressure / vac. service)

Std. Restrictor

R2324 Blue dot

Nut

N73

Ferrule

F73

Ferrule

F74

Both ferrules are required

CONVERSIONS:

1 PSI

=

2.04 inches of mercury (in. Hg.)